FL900 Logger with AV9000 Analyzer Module

  • Pembelian Minimum 1 Unit
  • Negara Asal American Samoa
  • KategoriAnalyzer Lainnya
  • Update Terakhir23 Mar 2020

Spesifikasi FL900 Logger with AV9000 Analyzer Module

FL900 Logger with AV9000 Analyzer Module


By pairing Hach's Sumberged AV Sensor with our state-of-the-art Hach AV9000 Analyzer Module, you create a powerful Hach Submerged Area Velocity Meter (FL900AV) which provides "cleaner", more precise data than ever before.

Our Submerged AV Sensor (sold separately) is a robust, 1 MHz Acoustic Doppler velocity sensor designed to measure wastewater flow with improved accuracy and reliability. This tried-and-true sensor also uses a pressure transducer to measure flow level and incorporates advanced technologies to ensure precision, including automatically correcting for temperature and velocity effects on measurements. With advanced signal processing and filtering options, the AV9000 Analyzer Module expands the applicability of the sensor into more difficult applications. Plus, advanced analyzer diagnostics, including capture and display of the Doppler spectra, allow you to verify that the sensor is working properly even before you leave the site, giving you peace of mind.

  • Reliable, Budget-Saving Monitoring
  • Increase Monitoring Crew Safety
  • Improved Accuracy
  • Easy Installation/Versatile Mounting Options

Analyzer Lainnya

Reliable, Budget-Saving Monitoring

The AV9000 Area Velocity Analyzer module is compensated
for temperature, thus eliminating potential velocity
errors of 2.7% over a 10ºC seasonal swing*. Its advanced
multi-scale digital Doppler analysis provides the optimal
combination of resolution and noise immunity. Mirror Image
Processing™ eliminates sign errors and the advanced Target
Set Processing™ reduces the impact of dominant targets
(particles) in the stream to deliver a more representative
velocity.

Increase Monitoring Crew Safety

With the time saving features designed into the FL900 Flow Loggers, crews spend less time in the manhole and less time on site to decrease monitoring costs while increasing the safety of flow monitoring crews.

Improved Accuracy

The AV9000 Area Velocity Analyzer module is compensated
for temperature, thus eliminating potential velocity
errors of 2.7% over a 10ºC seasonal swing*. Its advanced
multi-scale digital Doppler analysis provides the optimal
combination of resolution and noise immunity. Mirror Image
Processing eliminates sign errors and the advanced Target
Set Processing reduces the impact of dominant targets
(particles) in the stream to deliver a more representative
velocity.

Easy Installation/Versatile Mounting Options

The logger can be quickly attached to a wall, pole or manhole ladder in minutes. Users can choose to hang logger from standard carabiner or optional 4-bolt wall mount for pole, horizontal or vertical wall mount or ladder rung mount.

 Specifications

Accuracy:Timebase Accuracy: ± 0.002 % , synhchronized every 24 hours with server software and modem
Alarm:16 maximum channels
Allowable Level:Standard: 10.5 m (34.5 ft)

Extended: 31.5 m (103.5 ft)
Alternate KTO?:No
Battery Life:296 days With Sub AV Sensor with AV9000 Analyzer - 4 Lantern Batteries - (no modem, room temperature) at 15 minute logging intervals
Certifications:Logger: CE, optional AC power supply: UL/CSA/CE
Communication:Local Communications: USB, RS232 (Baud rates: 9600, 19200, 38400, 57600, 115200)
Communication (optional):Remote Communications (optional) Wireless modem: CDMA 1xRTT or GPRS
Datalog Channels:maximum
Data Storage:events maximum in non-volatile flash memory
Data Storage 2:Sample History: 2000 sample events maximum in non-volatile flash memory
Data Storage 3:Datalog: 325 k data points, 1128 days for 3 channels at 15-minute log intervals
Dimensions: Height:400 mm
Dimensions (H x W x D):400 mm x 254 mm x 220 mm
Environmental Rating:NEMA 6P
Level Accuracy:±0.16% full scale ±1.5% of reading
@ constant temp (±2.5ºC)

±0.20% full scale ±1.75% of reading
from 0-30ºC (32-86ºF)

±0.25% full scale ±2.1% of reading
from 0-70ºC (32-158ºF)
Level Measurement Method:Differential pressure transducer
with stainless steel diaphragm and
atmospheric pressure reference
Level Range:Standard: 0–3 m (0–10 ft)

Extended: 0–9 m (0–30 ft)
Logging:1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 or 60 minutes.
Manual Languages:English
Material:PC ABS structural foam
Modem:Non-Wireless
Operating Temperature Range:-18 - 60 °C at 95% Relative Humidity
Power Requirements (Voltage):8 - 18 V DC
Pre-configuration?:AV9000 Module
Protocol:Local Modbus RTU
Recommended Range:-1.52 to 6.10 m/s (-5 to 20 ft/sec)
Region:US
Sensor Connectors:1, no aux
Sensor Ports:1, 2 or 4 ports
Sensors Included:AV9000 Analyzer Module (Submerged AV Sensor sold separately)
Storage Conditions:-40 °C to 60 °C
Supported Sensors:Submerged Area Velocity Sensor
Typical Minimum Operating Depth:2 cm (0.8cm)
Velocity Accuracy:±2% of reading or 0.05 fps
Velocity-Induced Depth Error:Compensated based on flow
velocity
Velocity Measurement Method:Doppler ultrasonic;
twin 1 MHz piezoelectric crystals
Weight:Four batteries: 8.2 kg
Weight 2:No batteries: 4.5kg

Harga FL900 Logger with AV9000 Analyzer Module

: CALL

Dapatkan penawaran terbaik untuk
FL900 Logger with AV9000 Analyzer Module !

Nama
Quantity
Unit
Minta Penawaran

Produk Lainnya Dari PT ANDALAN BANGUN SEJAHTERA

IndoTrading.com adalah media iklan khusus Perusahaan, Toko, Distributor, Grosir, Agen, Supplier, Importir, Exportir, Trading dan Pabrik. Jika Anda tertarik untuk membeli FL900 Logger with AV9000 Analyzer Module ini dengan harga murah, Mohon hubungi Kontak Perusahaan ini diatas.

Jika Anda adalah perusahaan yang ingin menjual di IndoTrading.com , Anda bisa mendaftarkan perusahan Anda disini
Kontak Supplier
ANDALAN BANGUN SEJAHTERA

Verified Supplier
4 Tahun Perusahaan ini adalah Anggota Platinum Member Berbayar IndoTrading.com. Kami telah melakukan verifikasi Alamat dan Eksistensi Perusahaan ini.
Response Rate :
55%
Jl. Sumur Batu no. 1 Kemayoran Jakarta Pusat Jakarta Pusat, DKI Jakarta - Indonesia

Perusahaan Lainnya yang berkaitan dengan ANDALAN BANGUN SEJAHTERA

PT. Petronusa Timur Instrument

Verified Supplier
Indotrading gold member
2 Tahun
Response Rate 100%

CV. Kharisma Mandiri

Verified Supplier
Indotrading gold member
4 Tahun
Response Rate 100%

CV. Armoyo Kreasi Mandiri

Verified Supplier
Indotrading gold member
2 Tahun
Response Rate 100%
Arrow